Digestible and Resistant Starch Assay Kit K-DSTRS

Digestible and Resistant Starch determination

Description

The Digestible and Resistant Starch Assay Kit (K-DSTRS) for the determination of digestible, resistant and total starch in starch samples, plant and other materials.

This method is based on the research of Englyst et al. (Ref) with some modifications. Digestion is performed using saturating levels of pancreatic α-amylase (PAA) and amyloglucosidase (AMG), but in stirred containers rather than shaken tubes, to simplify sample removal.

In line with Englyst definitions:

  • Rapidly digestible starch (RDS) is that starch which is digested within 20 min.
  • Slowly digestible starch (SDS) is that starch which is digested between 20 and 120 min.
  • A new term, ‘Total digestible starch (TDS)’ is introduced (and measured) to cover all starch that is digested within 4 h (the average time of residence of food in the human small intestine).
  • Resistant starch (RS) then, is that starch which is not digested within 4 h.

Description

Content: 40 assays of each per kit
Shipping Temperature: Ambient
Storage Temperature: Short term stability: 2-8oC,
Long term stability: See individual component labels
Stability: > 2 years under recommended storage conditions
Analyte: Digestible Starch, Resistant Starch, Total Starch
Assay Format: Spectrophotometer
Detection Method: Absorbance
Wavelength (nm): 510
Signal Response: Increase
Linear Range: 4 to 100 μg of D-glucose per assay
Limit of Detection: 3.1 g/100 g
Reaction Time (min): ~ 360 min

Ref. K-DSTRS