BlueInOneFERM Gas Sensor O2 + CO2

Gas Sensor for O2 and CO2

Categories: ,

Description

  • parallel measurement of O2 and CO2
  • compact stainless steel housing
  • PAT conform in-situ- measurement
  • auto compensated humidity and pressure
  • no gas cooler, pumps or valves needed
  • connectable to any hose/tube or pipe
  • status display by luminous pushbuttons

Description of the O2 sensor

  • Measuring principle: Zirconium dioxide
  • Measuring range: 1 to 50 Vol.%
  • Accuracy: < 0,2% FS ± 3% value
  • Drift: < ± 2% value / year
  • Lifetime of sensor element: 15 000 operating hours
  • Temperature inside of the sensor unit: 580°C / 1076°F
  • Output: Slope 0.32 mA / Vol.%, 4 mA = 0.0 Vol.%, 20 mA = 50.0 Vol.%

Description of the CO2 Sensor

  • Measuring principle: Infrared, dual wavelength
  • Measuring range: 0 to 10 Vol.%
  • Accuracy: < 0.2% FS ± 3% value
  • Drift: < ± 2% value / year
  • Lifetime of optical components: approx. 3 years
  • Temperature inside of the sensor unit: 3 °C / 5.4 °F higher than max. temperature
  • Output: Slope 1.6 mA / Vol.%, 4mA = 0.0 Vol.%, 20mA = 10.0 Vol.%

Commun description

  • Temperature range: 15 – 40°C / 59°F – 104°F
  • Pressure range: 0.8 to 1,3 bar / 11.6 to 18.85 psi
  • Operating humidity: 0 to 100% RH
  • Housing: Stainless steel, IP65
  • Materials in contact with process gas: steel 1.4571, viton, sapphire, PTFE, Polymer H.L., nitrile, tygon
  • Filters: PTFE 0,22 um, PTFE 5 um
  • Power supply: 24V 1A
  • Don’t use in explosive atmosphere, in anoxic atmosphere, in gases with polymers or silicon components or in gases with halogens (F, Cl, Br), CFC or SOX and H2S