Description
Fully automated biochemistry analyser, 200 tests/hours, up to 280 tests/hour with the optional ISE. Test mode: end point, kinetics, fixed time, immunoturbidimetry,monochromatic, bichromatic, multistandard analysis, linear and non linear calibration, single and dual reagents. Completely open for editing test parameters and calculation parameters.
Sample System
- Sample tray: 44 sample position
- Sample volume:3 to 30ul, programmed by 0.1ul step
- Sample probe: Liquid level detection and Collision protection automatically
- Probe cleaning: automatic washing both inside and outside, Carry over<0.1%
- Automatic sample dilution
Reagent System
- Reagent tray: 44 positions for single &dual Reagents.
- Refrigeration: Refrigerated reagent, refrigerated compartment with Peltier element for 24h cooling system maintaining temperature between 4 and 15°C
- Reagent bottle size: 20ml and 35ml
- Reagent volume:150 to 300ul,programmed by 0.1ul step
- Reagent probe: Liquid level detection and collision protection automatically
- Pre-heating function inventory checking.
- Real-time monitoring of reagent volume and auto alarm
- Probe cleaning: automatic washing both inside and outside, Carry over<0.1%
Reaction System
- Reaction tray: 48 reaction cuvettes
- Reaction volume: 200 to 300ul
- Reaction Temperature:37°C+0.1°C
- Mixing system: Independent mixing probe
Optical System
- Light source: Halogen-tungsten lamp
- Wavelength: 340nm, 405nm, 450nm, 494nm, 505nm, 546nm, 578nm, 620nm, 670nm, 700nm
- Linear range:0 to 3ABS
Control and Calibration System
- Calibration mode: Linear(one point ,two point and multi-point),Logit-Log4P, Logit-Log 5P, spline, exponential ,polynomial, parabola
- Control rules: Westgard multi-rule
Operation Unit
- Operation system: Windows2000 system or above
- Interface: RS-232
- Power Supply: 220V+22V, 50/60HZ+1HZ2. Power consumption: 400VA
- Temperature:15 à 30°C
- Humidity: <=85%
- Water consumption: 5L/h
- Dimension: 760(L)X570(W)X1000MM
- Weight: approx.85kgs