Description
It can assemble two syringes, can be used as continuous transmission liquid syringe pump. Good performance, high precision and high accuracy. Durable mechanical transmission design, long life and good reliability. Support for multiple working mode. High resolution color touch screen LCD, simple and easy to set parameters, status display intuitive and clear;With adjustable protection agencies and alarm mechanism. The RS – 485 communication, support the MODBUS protocol, easy to connect to other devices;All stainless steel shell, easy to clean and maintain.
Flow: 0.184 nL/min-83.318 mL/min
Max: 2 Channels
Function and Feature of the syringe pump
• It can assemble two syringes
• A variety of working mode choice
• Color touch screen, LCD display, convenient and quick operation
• Support screen lock, key mute operation
• With lights, buttons, clearly working state
• Support a variety of standard syringe, customize syringe
• High precision control
• Syringe protection and traffic jam alarm
• RS485 communication, support the MODBUS protocol
• External signal control start-stop and direction
• A wide range of power input
• Full stainless steel shell
Technical parameters of the syringe pump
• Syringe size: 10 ul – 10 ml
• Work mode: infusion, withdraw, infusion/withdraw, withdraw/infusion, continuous mode
• Channel number: 2
• pusher advance per microstep: 0.156 um/ustep
• linear speed: 1 um/min – 150 mm/min
• Linear resolution: 1 um/min
• Linear travel accuracy: error <± 0.35% (>30% of pump stroke)
• Linear force: maximum rated: > 20 KGF
• Thrust regulation: 1-100% arbitrary adjustable
• Syringe selection: built-in main manufacturers, the main model syringe to choose;Can use custom syringe, direct input syringe size& diameter
• Flow calibration: through the calibration procedure to obtain more accurate fluid volume
• Display: 4.3 -inch color LCD, 480 * 272 resolution, color 65 k;With screen display and transfusion volume, quantity of liquid, flow, operation direction, syringe specification, etc.;Animation display operation status
• Operation mode: touch screen + button
• Power-off memory: storing the running parameters automatically
• Other features: pause and stop;The sound tip;The lock parameters;Fast forward and retreat quickly
• State signal output: 1 road stop state, 1 state road direction
• Control signal input: 1 way down along the priming, 1 way down along the trigger to stop
• communication interface: RS485, MODBUS protocol
• Applicable power: AC100-240 v / 20 w
• working temperature: 5 to 40 ℃
• Relative humidity: < 80% RH
• Dimension: 290 * 228 * 180 mm
• Weight: 4.8 kg
Syringe selection (Flow rate =Linear rate×Syringe inner sectional area)
Syringe size | Syringe ID(mm) | Min. flow rate (nl/min) | Max .flow rate(ml/min) |
10μl | 0.485 | 0.184 | 0.027 |
25μl | 0.729 | 0.417 | 0.0626 |
50μl | 1.03 | 0.833 | 0.125 |
100μl | 1.457 | 1.667 | 0.250 |
250μl | 2.304 | 4.169 | 0.625 |
500μl | 3.256 | 8.326 | 1.248 |
1ml | 4.699 | 17.342 | 2.601 |
5ml | 11.989 | 112.890 | 16.933 |
10ml | 14.427 | 163.169 | 24.754 |
20ml | 19.05 | 285.027 | 42.754 |
30ml | 21.59 | 366.090 | 54.913 |
60ml | 26.594 | 555.459 | 83.318 |