Description
it is is an automatic osmometer with a 40-place carrusel-based autosampler for the rapid determination of the osmolality in biological fluids, pharmaceutical samples and other solutions by means of the cryoscopic point variation.
Osmometer features
- Sample capacity: up to 40 samples
- Full colour 7” touchscreen display
- Bath with Peltier cells controlled by software
- Operating room temperature: +5°C to 36°C
- Automatic calibration
- Lash and agitation width controlled by software
- Sample volume: from 50 to 200 uL
- Test duration: about 2-3 minutes/sample (from 0 to 900 mOsm/Kg)
- Warm up time: about 5 minutes
- Analytical range: from 0 to 3 000 mOsm/Kg
- Resolution: 1 mOsm/Kg
- Repeatability and reproducibility: ±2.5 mOsm/Kg (from 0 to 600 mOsm/Kg) or ±0.5% of the value (from 600 to 3 000 mOsm/Kg)
- USB port for data downloading to a pen drive
- 2 x RS232 ports for connecting the provided thermal printer and an optional PC
- An additional RS232 port can be added for connecting an optional barcode reader (to be requested when ordering)
- Audit Trail function
- Users’ passwords obscured by means of “*****”
- The sample identification alphanumeric codes can reach up to 20 digits
- A second USB port allows the user to connect an optional barcode/QR-code reader for the quick identification of samples
- Easy updates of the firmware by means of a USB pen drive
- Possibility to use an optional plain paper printer instead of a thermal one
- Power absorption: max. 150 W
- Size: 330x610x360 mm (wxdxh, head down)
- Net weight: 20.5 Kg