Description
Its is a automatic freezing point osmometer is widely applicable in pharmaceutical enterprises, including manufacturers of injections, eye drops, and vaccines.
This osmometer features automatic sample injection, eliminating the need for a pipette. The sample volume is precisely controlled by the system, and sample detection is automatically performed. After the measurement is completed, the system automatically cleans the entire system, including the tubing and measurement components, without any human intervention.
Default calibration points are 2000, 700, and 100 mosm/kg, which can calibrate the measurement range from 0 to 2000 mosm/kg.
Custom calibration points: any two points other than 0 mosm can be calibrated, allowing users to customize the high calibration point and low calibration point according to their measurement requirements.
The measurement interface primarily presents the sample number, temperature curve, and measurement data in real-time, providing a clear and intuitive view of the testing process and data.
Technical Specifications
Sample Type: | aqueous solution |
Sample Volume: | 200μl or more |
Test Time: | Approx. 2 minutes/sample |
Sample Capacity: | 24 samples |
Resolution: | 1 mOsm/kg H2O |
Measurement Range: | 0-2000 mOsm/kg H2O |
Accuracy: | 0-400 mOsm/kg H2O: ±3 mOsm/kg H2O (1 SD) |
400-2000 mOsm/kg H2O: ±0.75% (1 SD) | |
Repeatability: | 0-400 mOsm/kg H2O: SD≤3 mOsm/kg H2O |
400-2000 mOsm/kg H2O: CV≤0.75% | |
Language: | Chinese, English |
Voltage: | 100-240 VAC(50/60Hz) |
Power: | 100W |
Dimensions (L×W×H): | 320x460x447 mm |
Weight: | 18 kg |
Storage Temperature: | -20℃~+45℃ |
Operating Conditions: | Temperature: 18°C ~ 35°C, 5% ~ 80% relative humidity (noncondensing) |
Delivered with the osmometer
- OM-6080 multi-sample osmometer
- Calibration Solution Set (incl. 700 and 100mosm calibration standard)
- Cleaning solution,500ml
- Cleaning bottle
- Waste bottle
- External dot matrix printer