Description
This mini spray atomizer is designed to ensure that all functions are simple to select and adjust, so that optimal conditions for spray drying can be achieved quickly.
Since the sample is dried directly from the solvent and suspension into a micronized powder, there are fewer pre-treatment and post-treatment operations, such as filtration, separation and pulverization,that are traditionally associated with powder production, and contamination during these operations is avoided.
Advantages
- By connecting the organic solvent recovery unit GAS411C, samples containing organic solvents can be handled professionally and safely.
- The machine is equipped with a motorized lift for easy installation and dismantling of the ancillary equipment.
- The machine is equipped with a power outlet (2A) for supplying power to the stirrer to facilitate sample feeding while stirring the suspension.
- he unique peristaltic pump, nozzle cooling mechanism, pulsed nozzle cleaning mechanism, and anticlogging stabbing needle are used to achieve a wide variety of spray conditions and stability.
Features
- Water evaporation Max.1 500mL/h
- Temperature adjustment range: 4to 240 °C (inlet temp.), 0 to 100 °C (outlet temp.)
- Temperature accuracy: ±1 °C
- Adjusting range for drying air: 0 to1.0m3/min
- Spray air flow adjustment range: 0 to 30L/min
- Nozzle air pressure adjustment range: 0 to 0.6MPa (adjustable)
- Nozzle washing function: Auto clean out needle, pulse jet cleaning
- Temperature output for Inlet temp. and Outlet temp. : 4~20mA
- Temperature control: Multi-stage PID control
- Touch screen PID digital temperature adjusting device
- Touch panel: Temperature adjustment, blower, heater, liquid feed pump, switch for pulse jet, automatic needle poking, alarm display, operation
- Control select switch: select inlet temperature, outlet temperature control switching
- Temperature sensor : PT100
- Heater : 3.2KW
- Sample feed pump One touch pedestal type
- Spraying from Nozzle: Use air compressor for spraying (sold separately) or use GAS411C built-in compressor (N2) when connected to organic solvent recovery device GAS411C (sold separately)
- Service outlet For stiller etc.:AC220V, 2A
- Blower : Tube blowers (can be switched to suction or push-out)
- Filter: Suction filter, exhaust filter
- Solvent recovery: Use of solvent recovery unit GAS411C (sold separately)
- Spray nozzle cooling structure: Fitting×2, outer diameter φ10.5mm (optional cooling water circulation device CF312L-B)
- Air connection for spraying : Outside diameter of connector, φ7mm
- Spraying air pressure: 0.3MPa
- Exhaust connection : Diameter 50mm
- Safety functions :Inlet and outlet temperature overheating, Pump reversal function, Overcurrent leakage protection switch, Nozzle connection abnormality (when connected to GAS411C)
- External dimensions : W760×D420×H1350mm
- Weight : 110kg
- Power supply AC200~230V 50/60Hz, 16~18A
- Accessories: 2 fluid supply hoses, 1 exhaust hose (with 1 hose tie), Exhaust adapter, Outlet temperature sensor, fuse (250V 2A), De-static cable, Inlet hose 5m (with 2 hose ties), Nozzle adapter bushing, Stand assembly, Protective cover (COV30)
- Option device: GWS410 (exhaust gas cleaning equipment), External filter, Dehumidification equipment, USB data storage, Various nozzle sizes