Description
This device allows to calibrate humidity probes with excellent precision. It does not regulate temperature but only humidity. It has a temperature probe for measurement and a HC2A-S probe for humidity. It allows to calibrate up to 7 humidity probes.
Advantages
- Portable device with optional battery and hard carrying case
- Intuitive interface facilitates automation of probe verification
- Simultaneous validation of 7 probes
- Automated validation procedures for complete probe verification
- Internal calibration correction cycle ensures continued confidence
- Optional chilled mirror reference hygrometer, allowing it to be used as a calibration
- Download recorded verification data directly to USB memory
Features
| Humidity range | 5 to 95 % RH |
| Humidity stability | +/-0.5 % |
| Humidity uniformity | +/-0.5 % |
| Stabilisation time | Typically <5 min |
| Humidity accuracy | +/- 0.8 % |
| Temperature accuracy | +/- 0.1 °C |
| Display | color LCD 4.3” |
| Sensors interface | 7×24 V, accepted signals 0 to 20 mA, 4 to 20 mA, 0 to 1 V, 0 to 5 V, 0 to 10 V |
| Resolution on the display | 0.1 |
| Data recording | Internal memory 2 Go |
| Battery | 1500 mAh |
| Power supply | 24 VDC (100 to 240 V AC with the included adapter) |
| Sensor diameter | 12 mm, 13.5 mm, 14 mm, 15 mm, 18.5 mm, 19 mm, 24 mm, 25 mm |
| Chamber volume | 1050 cm3 |
| maximal deepth | 60 mm |
| Water capacity | 25 cm3 |
| Desiccant capacity | 25 cm3 |
| Operating conditions | 5 to 40 °C |
| Dimensions | 100x250x300 mm |
| Weight | 3.2 kg |