Description
Equipped with a xenon flash lamp and monochromator-based optics, this microplate reader delivers full-spectrum measurements across the 190–1,100 nm wavelength range without requiring optical filters.
The MRX Series features a monochromator-based optical system with photodiode detection, enabling wavelength selection in 1 nm increments and simultaneous measurement at up to eight wavelengths.
Designed for research, quality control (QC), and educational laboratories, it supports a wide range of absorbance-based assays with broad plate compatibility (6 to 384 well plate) and flexible measurement capabilities.
The MRX View PC software enables reliable execution of various experiments. It enhances usability and application flexibility through a wide range of features, including control over temperature, shaking, measurement area, and lag time settings.
The microplate reader is available in 3 different models.
- MRX A2000 (MRC): Shaking, Incubating, Cuvette measurement
- MRX A2000 (MR): Shaking, Incubating
- MRX B2000: Shaking
Features
Detection modes: UV-Vis absorbance
Light source: Xenon flash lamp
Detector: Photodiode
Wavelength selection: Monochromator
Wavelength range: 190 to 1,100 nm / 1 nm increments
Bandwidth: 2.9 nm
OD Linearity: < 1% from 0 to 2.5 OD
OD Accuracy: < 1% at 2.0 OD / < 3% at 2.5 OD
OD Repeatability: < 0.5% at 2.0 OD
Read-out range: 0 to 4.0 OD
Resolution: 0.0001 OD
Pathlength correction: yes
Wavelength accuracy: ± 1.5 nm
Wavelength repeatability: ± 0.2 nm
Stray light: 0.03 % at 230 nm
Measurement speed (from A1 back to A1): 8 seconds with 96-well plate
Shaking: Linear, orbital, double orbital
Read methods: Endpoint, Kinetic, Spectral scanning, Well area scanning
Microplate types: 6-, 96-, 384-well plates
Incubation range: From ambient + 2 °C to 65 °C
User interfaces: PC software
Connections: Ethernet port
Power: 110/220 V, 50/60 Hz
Power Consumption: 100 W
Weight: 12 kg
Dimensions(W*D*H): 340 mm * 410 mm * 225 mm