Description
This paramagnetic oxygen sensor module is designed for incorporation in your analyzer systems or for the solution of oxygen measuring problems using a microprocessor or a PLC-based system.
The sensor module is maintenance-free, have a long lifetime, give a rapid and accurate signal response and is virtually insensitive to other gases. It has a compact design which meets high quality standards and easy implementation.
Technical Specifications
- Standard measuring ranges 0 – 25% O2, 0 ‐ 100% O2, other ranges on request
- Minimum measuring range 0 ‐ 5% O2
- Output value standard 4 ‐ 20 mA (galvanic isolated)
- Output value (option)0 ‐1V, 0 ‐ 4V (galvanic isolated)
- Flow: 18 ‐ 70 l/h with fix Bypass (Standard), Maximum 250 ml/min (without Bypass)
- Operating gas pressure: +/-300 hPa (0.3 bar) in operation, +/- max. 1000 hPa (1bar)
- Operating temperature : 5 – 45°C, (heated measuring cell = 55°C)
- Storage and transport temperature: -25°C to +65°C
- Relative humidity: 0 – 90% r.h.
- Dimension (W x H x D): 80x78x93 mm without electrical connector gas connection
- Weight ca. 1090 g
- Repeatability < ±0.03% O2 [time base for gas switch >= 5 min]
- Zero point drift < ±0.1% O2 / week [may be higher during the first days after putting into operation or after longer period of storage or transport]
- Temperature influence at zero < ±0.05% O2 /°C
- Flow error < 0.2% O2 for increase of flow from 20 to 100 ml/min
reduction to < 0.1% with build-in fix bypass [optional] - T90-time < 3s with 150 ml/min flow and gas change from nitrogen to air
- Voltage 12-28 VDC, Power consumption 12 W