Ultrasonic Flaw Detector NOVOTEST UD2301

Ultrasonic flaw detector, material thickness measurement.

Categories: ,

Description

The Ultrasonic flaw detector UD2301 is designed for non destructive testing of metals, plastics, glass, composite materials, weld inspection and measurement of the thickness of the various products and structures.

The Ultrasonic Flaw detector is made in small enclosure with the optimal size for performing control in tight spaces. Device equipped with a clear color display with a high resolution of 320 * 480 pixels, which significantly improves the usability of the device.

Appointment of flaw detector

  • Control of the welds and base metal
  • Places of corrosion, cracks, delaminations and other internal defects
  • Determination of the coordinates and parameter estimation of defects such as discontinuity and
    uniformity of material in products from metals, plastics, composites and other materials
  • Measurement of the thickness of the product.

The Flaw detector allows to solve a wide range of tasks – from the thickness measurement of thin
products to large-sized casting flaw detection.

  • Operating Frequency Range: from 1 to 10.0 MHz
  • The range of measured time intervals (duration of scanning): from 6 to 1000 μs
  • Velocity range: 1000 – 9999 m / s
  • The error of measurement of time intervals: not exceed ± 0.025 μs
  • Maximum permissible error of measurement of the amplitudes of the signals at the receiver input in the range from 0 to 110 dB: not exceed ± 0.5 dB
  • Testing gain range: 125 dB
  • Averaging over the quantity of starts: from 1 to 16
  • The range of variation of temporal sensitivity adjustment (TVG): 40 dB
  • Number of control points TVG: 15
  • The duration of the excitation pulse to the load: from 0.0 to 0.5 μs
  • The amplitude of the excitation pulse to the 50 Ohm load, not less: 100, 200, 300
  • Operating frequency range of the receiver on the level -3 dB: from 1 to 10 MHz
  • Deviation of the amplitudes of input signals in the range from 10 to 100% of the screen height not more than 1 dB
  • Scanning: from 1 to 1000 μs
  • Delay of scanning: from 0 to 2000 μs
  • Range of measurement of time intervals: from 0 to 1000 μs
  • Setting a delay in the prism of the probe: from 0 to 15 μs
  • Automatic signaling of defects (AFS): dual-gate
  • Setting range of AFS gates: from 0 to 2000 μs
  • Adjusting thresholds of AFS gates: from 0 to 100 % of display height
  • Detection of signals: positive half-wave, radio mode
  • Dimensions (WxHxD): 80х162х38 mm;
  • Weight, not more: 250 g (without batteries)
  • Mean time between failures: not less than – 3000 hours

Standard set of device:

  • Electronic unit of Ultrasonic Flaw NOVOTEST UD2301
  • Probes 2 pcs
  • Cable Lemo-Lemo 1 pc
  • AA batteries – 3 pcs, charger
  • USB cable for PC connection
  • Operating manual
  • Case

Additional information

Model

TB 1000-0.1F, TB 1000-0.1N, TB 1000-0.1FN, TB 2000-0.1F