Description
The system allows the extraction of DNA and RNA from any type of sample using magnetic microbeads. The extraction process is fully automated.
After enzymatic digestion and solubilization of the lipids in a special solution, the negatively charged DNA is brought into contact with magnetic beads whose surface is positively charged. DNA will therefore cling to the surface of these beads. The solution is removed while the DNA remains attached to the beads held by a magnet. The beads are then put back into another solution in order to neutralize their charges and the DNA is then released into the liquid.
The device allows you to program each step of the extraction.
For example: fixation of magnetic beads, lysis, wash 1, wash 2, elution, deposition of magnetic beads
Ready-to-use kits can be offered.
Application: PCR tests
Parameters
- Sample Volume: 20-500μL
- Sample Throughput: 1-32
- Magnetic Bead Recovery: >98%
- Extracting the Difference Between Holes: CV≤3%
- Heating Temperature: 8 independent heating modules, customize lysis and elution temperature (temperature range) according to your needs
- Oscillating Mixing: Low,medium and high third gears are adjustable, and the fluctuation range can be adjusted with the reagent volume
- Reagent Type: Magnetic bead open platform
- Extraction Time: 15-40 min/round (depending on the reagent used)
- Internal Program: Be able to store >100 groups of programs
- Program Management: Powerful program editing capabilities to meet different reagent needs. U disk program import and export can be achieved
- Safety Door Design: After the safety door is opened,the program operation will be automatically suspended, and the program can continue t run after the safety door closed
- Built-in Air Duct: No
- Ultraviolet Irradiation: Yes
- Packing Size(W*D*H)mm: 580*510*700
- Gross Weight(kg): 51